microscope:sem:pgt
Differences
This shows you the differences between two versions of the page.
Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
microscope:sem:pgt [2014/06/26 10:18] – mcmaster | microscope:sem:pgt [2016/11/15 07:43] (current) – removed mcmaster | ||
---|---|---|---|
Line 1: | Line 1: | ||
- | Princeton Gamma-Tech (PGT, now Thermo Scientific). | ||
- | |||
- | "X-Ray and Nuclear products have been discontinued and no longer offered for sale by Thermo Fisher Scientific. For Service and Support...Thermo Fisher Scientific has entered into an agreement with Berkeley Nucleonics Corporation (BNC) to provide warranty and non-warranty support for X-Ray and Nuclear products" | ||
- | |||
- | However, BNC documents seem to pitch it as part of their company and I've seen "BNC Princeton Division" | ||
- | |||
- | x-ray product summary: | ||
- | |||
- | ^ Image ^ Model ^ Type ^ Description ^ Notes ^ | ||
- | | [{{: | ||
- | | | 483 Controller | | cryocooler controller| | ||
- | | | PRISM 60 series | | EDS detectors | | | ||
- | | [{{: | ||
- | | {{jm: | ||
- | | | LS10138 | EDS detector (Si(Li)) / preamp | | [BNC Ge] | | ||
- | | | LS10143 | EDS detector (Si(Li)) / preamp | | [BNC Ge] | | ||
- | | | LS30143 | EDS detector (Si(Li)) / preamp | | [BNC Ge] | | ||
- | | [{{: | ||
- | | | LS60158 | EDS detector (Si(Li)) / preamp | | [BNC Ge] | | ||
- | | | LS60163 | EDS detector (Si(Li)) / preamp | | [BNC Ge] | | ||
- | | | IGX10125 | EDS detector / preamp | | [BNC Ge] | | ||
- | | | IGX10138 | EDS detector / preamp | | [BNC Ge] | | ||
- | | | IGX30133 | EDS detector / preamp | | [BNC Ge] | | ||
- | | | IGX30138 | EDS detector / preamp | | [BNC Ge] | | ||
- | | [{{: | ||
- | | [{{: | ||
- | | | RG-11 | Preamp | | Use w/ IGC/NIGC detector [NPC] | | ||
- | | | PO-14B | Preamp | | Use w/ NIGC detector[NPC] | | ||
- | | | RG-11 | Preamp | | Use w/ IGP detector [NPC] | | ||
- | | | TR-14 | Preamp | | Use w/ IGP detector [NPC] | | ||
- | | | TR-1 | Preamp | | Use w/ Si(Li)\IGX detector [NPC] | | ||
- | | [{{: | ||
- | | | MCA2100R-EN | MCA (accessory? | ||
- | | [{{: | ||
- | | [{{: | ||
- | | [{{: | ||
- | | [{{: | ||
- | | [{{: | ||
- | | [{{: | ||
- | | [{{: | ||
- | | [{{: | ||
- | | [{{: | ||
- | | [{{: | ||
- | | | QuantumMCA | Software | Quantitative | | | ||
- | | [{{: | ||
- | | | Quantum GeD | Software | quantium + qualitative (Ge detector only) | | | ||
- | | | Quantum Ge | Software | Quantum GeD minus deconvolution analysis | | | ||
- | | | Quantum NaID | Software | quantium + qualitative (NaI detector only) | | | ||
- | | | Quantum NaI | Software | Quantum NaID minus deconvolution analysis | | | ||
- | | | | | | | | ||
- | |||
- | Where: | ||
- | * MCA: Multichannel analyzer | ||
- | |||
- | |||
- | ====== PRISIM 60 series ====== | ||
- | |||
- | older units | ||
- | |||
- | |||
- | ====== PRISIM 2000 series ====== | ||
- | |||
- | Official title: " | ||
- | |||
- | Detectors [manual]: | ||
- | |||
- | Window type: ultra thin | ||
- | |||
- | Cooling method: LN2 or JT Cool | ||
- | |||
- | ^ Model ^ Crystal ^ Crystal size (mm*2) ^ FWHM (eV) ^ | ||
- | | PRISIM 2000 | Si(Li) | 10 | 129 | | ||
- | | PRISIM 2000 | Si(Li) | 30 | 135 | | ||
- | | PRISIM 2000 | Si(Li) | 60 | 148 | | ||
- | | PRISIM 2000 | HpGe | 10 | 115 | | ||
- | | PRISIM 2000 | HpGe | 50 | 129 | | ||
- | |||
- | |||
- | Typical system [Gammy and X-Ray detectors]: | ||
- | * preamp | ||
- | * PGT Model 326 LN2 level detector | ||
- | * PGT Model 312, 314 or 315 HV PS | ||
- | * Pulser | ||
- | * PGT Model 347 Amplifier | ||
- | * ADC | ||
- | * MCA | ||
- | |||
- | |||
- | ===== Pre-amp ===== | ||
- | |||
- | Connections: | ||
- | |||
- | ^ Label ^ Type ^ Notes ^ | ||
- | | TEST | BNC | | | ||
- | | BIAS | RP BNC / HV thing? | | | ||
- | | POWER | DB-9M | | | ||
- | | OUTPUT | BNC | | | ||
- | | INHIBIT | BNC | | | ||
- | |||
- | TODO: get better info on function/ | ||
- | |||
- | |||
- | ===== JM unit ===== | ||
- | |||
- | {{jm: | ||
- | |||
- | This unit was bought for reasonably low cost with hope to learn about them and maybe fix it. Worst case, this unit should work under atmosphere or scrap for the dewar | ||
- | |||
- | |||
- | ==== Instrument ==== | ||
- | |||
- | Front: | ||
- | |||
- | {{jm: | ||
- | {{jm: | ||
- | |||
- | Sides: | ||
- | |||
- | {{jm: | ||
- | {{jm: | ||
- | |||
- | |||
- | Vacuum section: | ||
- | |||
- | {{jm: | ||
- | {{jm: | ||
- | |||
- | Close up of broken window: | ||
- | |||
- | {{jm: | ||
- | |||
- | The screen is still intact. | ||
- | |||
- | {{jm: | ||
- | |||
- | My guess is that it got damaged during cleaning | ||
- | |||
- | Top: | ||
- | |||
- | {{jm: | ||
- | |||
- | Close up of cryocooled electronics (FET + pre-amp): | ||
- | |||
- | {{jm: | ||
- | |||
- | Covered removed: | ||
- | |||
- | {{jm: | ||
- | {{jm: | ||
- | |||
- | |||
- | ==== Tank ==== | ||
- | |||
- | Labels: | ||
- | |||
- | {{jm: | ||
- | {{jm: | ||
- | |||
- | Reads: | ||
- | MODEL 0PJ013-1038 | ||
- | SERIAL 2051 | ||
- | BIAS -600 VDC | ||
- | MADE IN USA | ||
- | PRINCETON GAMMA-TECH | ||
- | |||
- | Instrument mating flange (connected): | ||
- | |||
- | {{jm: | ||
- | {{jm: | ||
- | |||
- | Instrument mating flange (disconnected): | ||
- | |||
- | {{jm: | ||
- | |||
- | Tank side thermal conduction path: | ||
- | |||
- | {{jm: | ||
- | |||
- | Looks to be some sort of braided copper alloy | ||
- | |||
- | |||
- | ====== XS169-IXI ====== | ||
- | |||
- | [{{: | ||
- | |||
- | From listing: "Item removed from a Cambridge Scanning Electron microscope (I believe an s200)" | ||
- | |||
- | Preamp: | ||
- | |||
- | [{{: | ||
- | [{{: | ||
- | |||
- | Tank: | ||
- | |||
- | [{{: | ||
- | |||
- | Tank label: | ||
- | |||
- | [{{: | ||
- | |||
- | Reads: | ||
- | MODEL XS169-IXI | ||
- | SERIAL 1321 | ||
- | BIAS -600 VDC | ||
- | MADE IN USA | ||
- | PRINCETON GAMMA-TECH | ||
- | ... | ||
- | |||
- | |||
- | ====== LS30148 detector ====== | ||
- | |||
- | [{{: | ||
- | [{{: | ||
- | |||
- | This unit appears to have ConFlat connections and a fixed angle detector. | ||
- | |||
- | |||
- | ====== JT Cool ====== | ||
- | |||
- | [{{: | ||
- | |||
- | Small crycooler to replace the LN2. Requires Model 483 Controller (white box thing). | ||
- | |||
- | Summary [NPC]: | ||
- | * Uses " | ||
- | * Claims 5 years of maintenance free operation | ||
- | * Supports up to 50 foot gas lines | ||
- | * " | ||
- | * 10-35C environment | ||
- | * Down to 70K | ||
- | * Copper gas line option | ||
- | |||
- | |||
- | ===== Pix ===== | ||
- | |||
- | [{{: | ||
- | |||
- | [{{: | ||
- | |||
- | |||
- | ====== Imagemaster ====== | ||
- | |||
- | [{{: | ||
- | [{{: | ||
- | |||
- | They had a LS30148 and a PRISM 60 for sale along with it, unclear if they somehow go together. | ||
- | |||
- | |||
- | ====== IGC10-ES ====== | ||
- | |||
- | Overview: | ||
- | |||
- | | [{{: | ||
- | |||
- | Detector area: | ||
- | |||
- | | [{{: | ||
- | |||
- | Label: | ||
- | |||
- | | [{{: | ||
- | |||
- | Reads: | ||
- | MODEL IGC10 ES | ||
- | SERIAL 1502 | ||
- | BIAS + 2500V | ||
- | |||
- | |||
- | ====== References ====== | ||
- | |||
- | * Nuclear products catalog [NPC]: http:// | ||
- | * PRISM 2000 Si(Li) and HpGe DETECTOR SERIES USERS MANUAL ("2000 manual" | ||
- | * Princeton Gamma Tech " | ||
- | * NUCLEAR PRESENTATION [Nuclear presentation]: | ||
- | * BNC GERMANIUM AND SILICON X-RAY DETECTORS [BNC Ge]: http:// | ||
- | * eBay 191140107113: | ||
- | * eBay 400720324518: | ||
- | * eBay 270642145582: | ||
- | * eBay 360526462914: | ||