| VOLTAGE CONTROL | ±3.5 kV Configuration | ±7 kV Configuration |
| | |
| Range | 100-3500 V | 200-7000 V |
| CAMAC Programming | 14 bits | 14 bits |
| CAMAC Step | 0.25 V | 0.5 V |
| CAMAC Accuracy | ² ±(1.5 V + 0.1%) | ² ±(3 V + 0.1%) |
| Front-Panel Adjustment | Ten-turn vernier potentiometer |
| 500 V/turn | 1000 V/turn |
| CAMAC Monitor | 12-bit ADC | 12-bit ADC |
| CAMAC Resolution | 1 V | 2 V |
| CAMAC Offset (response to 0 V output) | 0-2 counts | 0-2 counts |
| CAMAC Accuracy | ±0.2% | ±0.2% |
| Front-Panel Monitor | Low Impedance (< 10 ohm) voltage output, suitable for driving a meter; BNC connector. |
| Front-Panel Scale | 2 V/kV ±0.5% | 1 V/kV ±0.5% |
| Front-Panel Offset (value for 0 V output) | ² ±4 V referred to HV output | ² ±8 V referred to HV output |
| | |
| CURRENT CONTROL | | |
| | |
| Range | 0-2.5 mA | 0-1 mA |
| CAMAC Programming | 8 bits | 8 bits |
| CAMAC Step | 10 µA | 4 µA |
| CAMAC Accuracy | ±(1% + 12 µA) | ±(1% + 5 µA) |
| Front-Panel Adjustment | Ten-turn vernier potentiometer |
| 250 µA/turn | 100 µA/turn |
| CAMAC Monitor | 12-bit ADC | 12-bit ADC |
| CAMAC Resolution | 625 nA | 250 nA |
| CAMAC Offset (Response to 0 µA load) | 1 or 2 counts | 1 or 2 counts |
| CAMAC Accuracy | ±0.5% | ±0.5% |
| Front-Panel Monitor | Low Impedance (< 10 ohm) voltage output, suitable for driving a meter; BNC connector. |
| Front-Panel Scale | 4 V/mA ±0.5% | 10 V/mA ±0.5% |
| Front-Panel Offset (value for 0 V output) | ² ±4 µA | ² ±2 µA |
| | |
| GENERAL | | |
| | |
| Output Power Rating | 7 W | 7 W |
| Output Ripple (at maximum current) | < 50 mV rms | < 50 mV rms |
| Output Temperature Coefficient | Typically ±50 ppm/¡C, ±100 ppm/¡C maximum |
| Voltage Regulation | < 0.5 V (0 to 2.5 mA) | < 1 V (0 to 1 mA) |
| Short Circuit Protection | Yes | Yes |
| Output Connector | SHV | Reynolds 1064-1 |
| | |
| CONTROLS | | |
| | |
| HV On/Off | Toggle switch; turns On/Off HV output. |
| Manual/Remote | Selects front panel or CAMAC operation of voltage and current set. |
| ADC | 12 bits; conversion time < 30 µsec. |
| | |
| INDICATOR LAMPS | | |
| | |
| HV On | Voltage present at output. |
| Overload | Supply in current-limiting mode. |
| Positive | Positive output configuration. |
| Negative | Negative output configuration. |
| 3.5 kV | ON | OFF |
| 7 kV | OFF | ON |
| | |
| POWER | 350 mA at +24 V; 350 mA at -24 V; 350 mA at +6 V. |