+ | ====== Operation ====== | ||
+ | |||
+ | 2018-04-24 notes | ||
+ | * Summary: generally seems to work, reliability issues possibly software related | ||
+ | * Booting | ||
+ | * Embedded controller ethernet must have system (ie my laptop) at other end at boot | ||
+ | * Embedded controller stage connection must be attached at boot | ||
+ | * Cannot be activated without restarting system | ||
+ | * GUI will complain and then segfault, but otherwise seems operational | ||
+ | * GUI | ||
+ | * Sometimes signal disappears and can be made to come back by switching signal views around (ie left pane vs right pane) | ||
+ | * Sometimes GUI becomes completely unresponsive and has to be rebooted. But seems to only happen when changing settings, ie not during imaging cycles | ||
+ | * Pressure | ||
+ | * Around 2E-5 typical pressure after pumping down | ||
+ | * BSE not working | ||
+ | * Wiggling power cable RJ-11 causes noise on screen | ||
+ | * Removing signal BNC results in blank screen | ||
+ | * Next | ||
+ | * Check power supply | ||
+ | * Scope on BNC | ||
+ | * EDS still not tested | ||
+ | * Adjusting XY centering helped bring up beam energy | ||
+ | * Current was something like 25 uA...think target is 75 uA => probably need to adjust something | ||
+ | * Was able to select different acceleration voltages | ||
+ | * Focusing | ||
+ | * Set image to 256x256 | ||
+ | * Set scan speed relatively high | ||
+ | * Adjust slider until in focus | ||
+ | * Increase resolution if need to get finer focus | ||
+ | * Didn't try using auto-focus this time | ||
+ | * Degaussing may bring out of focus | ||
+ | * Changing accelerating voltage brought out of focus slightly maybe | ||
+ | * Higher acceleration voltages are fuzzier (makes sense => deeper penetration => more scattering) | ||
Open questions / TODO: | Open questions / TODO: |